Page 305 - Book Hosokawa Nanoparticle Technology Handbook
P. 305

FUNDAMENTALS                            CH. 5 CHARACTERIZATION METHODS FOR NANOSTRUCTURE OF MATERIALS
                    TEM provides a phase contrast imaging of a thin  5.3.1.1 Interatomic force
                  bulk. However,  AFM provides real topographical  Interatomic force between two electrically neuter
                  image of sample surface. A typical vertical distance  atoms is described by Lennard–Jones equation [1].
                  resolution is very high with 0.01 nm, which is much
                  better than normal SEM and TEM. A scanning area of                ⎧ ⎛ ⎪  
⎞  12  ⎛  
⎞ ⎪ ⎫
                                                                                              6
                  typical AFM is from several nanometers to several         Ur()   4  ⎨ ⎜ ⎟    ⎜ ⎟ ⎬   (5.3.1)
                  hundreds nanometers. The maximum height for imag-                 ⎩ ⎪ ⎝  r ⎠  ⎝  r ⎠  ⎭ ⎪
                  ing also decreases with a decrease in scan area.
                    Fig. 5.3.1 shows each imaging size of SPM and  where, 
 and   are interatomic distance and energy in
                  other typical microscopes. AFM can work perfectly  balance conditions, respectively. Both 
 and   are con-
                  well with atomic resolution in vacuum, ambient air,  stant values determined by molecular species.  r is
                  or even a liquid environment. AFM is used not only  interatomic distance. Fig. 5.3.2 shows the interatomic
                  for surface observation but also for microfabrication  force versus interatomic distance.  The interatomic
                  or measurement of surface physicality. AFM is use-  force at large distance shows weak attraction due to
                  ful for imaging of both conducting sample and insu-  induced dipole moment (dispersion interactions).
                  lating one, e.g., polymer material, biological  However, the force at small distance shows repulsion
                  macromolecules, chemically modified nanoparti-  due to exchange interaction from Pauli exclusion
                  cles. Comparison of AFM and other microscopes is  principle. The interatomic distance,  r at the lowest
                                                                                               1
                  shown in Table 5.3.1.                          force shown in Fig. 5.3.2 corresponds to the inter-
                                                                 atomic distance in closed-packing.
                                                                   5.3.1.2 Detection technique of force
                    1 mm                                         The AFM consists of a sharp tip (probe) with a tip
                                 Scanning Electron Microscope
                                 (SEM)                           radius of curvature on the order of nanometers at the
                                                                 end of a leaf spring or “cantilever”. The cantilever is
                                                                 microfabricated from silicon or silicon nitride.
                                                                 Several manufacturers produce many cantilevers with
                    z measurement range  1  m  Transmission Electron Microscope [TEM]  and tip radius of curvature. The cantilever selection
                                                                 various geometries, various values of spring constant,
                                                                 for each sample is one of important factors to obtain
                                           Optical Microscope
                                                                 clear and reproducible images.
                                                                  As the tip of the cantilever approaches the sample
                                                                 induces the vertical deflection of the cantilever. AFM
                     1 nm                                        surface, the force between the tip and sample surface
                                                                 obtains the force to detect the deflection of the can-
                                                                 tilever. There are two optical methods to detect the
                                Scanning Probe Microscope (STM, AFM)  deflection of the cantilever, optical lever method, and
                                                                 laser interferometry.
                                1 nm        1  m        1 mm      Fig. 5.3.3a shows the concept of the optical lever
                                    x-y measurement range        method used mostly. The sample is located on the
                                                                 tube piezo scanner, which moves the sample in the
                  Figure 5.3.1                                   x–y–z directions (In some AFMs, the piezo scanner
                  Each imaging size of SPM and other typical microscopes.  is mounted to the cantilever.).  The four-segment



                  Table 5.3.1
                  Comparison of AFM and other microscopes.

                                      AFM                 TEM                 SEM            Optical microscope
                  Maximum        Atomic resolution   Atomic resolution   Several nanometers  Several hundreds
                  resolution                                                                 of nanometers
                  Observation    In air, in liquid,
                  environment    in vacuum, in gas   In vacuum           In vacuum           In air, in liquid
                  In situ        Possible            Impossible          Impossible          Possible
                  observation
                  Preparation    Easy                Difficult           Easy                Easy
                  of sample


                  280
   300   301   302   303   304   305   306   307   308   309   310