Page 350 - Biomedical Engineering and Design Handbook Volume 2, Applications
P. 350

328  DIAGNOSTIC EQUIPMENT DESIGN

                       but the easiest to implement is the subtraction method where information is simultaneously acquired
                       into a second energy window centered below the photopeak in Compton scatter region of the energy
                       spectrum. After establishing an appropriate normalization factor, the counts from the scatter window
                       are subtracted from the photopeak window and the corrected projections are then used in the recon-
                       struction algorithm.
                         One other correction that has been implemented with SPECT studies is the compensation for
                                   14
                       spatial resolution. As discussed in the section on scintillation cameras, the spatial resolution depends
                       on the source to collimator distance. As a result this correction cannot be made with the analytic recon-
                       struction methods (i.e., filtered backprojection) but has been implemented with iterative reconstruction
                       algorithms.


           11.3.1 SPECT Image Reconstruction
                       The details of SPECT image reconstruction are beyond the scope of this article, but the interested
                       reader can see the details in the cited literature. 15,16  Because SPECT image sets are relatively small
                       compared to other medical imaging modalities, the computational and display requirements can be met
                       by personal computers. However, the integration of SPECT studies with CT and MRI put increased
                       demands on memory and storage.


           11.3.2 SPECT System Performance
                                                                     4
                       Typical performance specifications for SPECT imaging systems are summarized in Table 11.3. As
                       with conventional planar imaging, the scintillation cameras and the associated collimation are the
                       primary factors affecting the performance. SPECT spatial resolution is nearly isotropic with a
                       FWHM of 8 to 10 mm for brain imaging where the detectors can get close to the radioactive source.
                       The spatial resolution degrades to 12 to 18 mm for body imaging because the detectors cannot be
                       positioned as close. The components of SPECT spatial resolution and their relative importance can
                       be identified from the equation shown below:

                                                          2
                                                R SPECT  =  R col  +  R 2 filter  +  R 2 int
                       As before, R int  and R col  represent the intrinsic and collimator resolution components. R filter  is the
                       FWHM of the smoothing kernel required to yield an acceptable reconstruction. The intrinsic spatial
                       resolution is the least important factor in this calculation since it is usually a factor of 2 or more
                       smaller than the other components. The trade-off between spatial resolution and count sensitivity is
                       explicit in this equation. Decreasing R col  to improve spatial resolution will often require R filter  to
                       become larger to compensate for increased noise.



                                    TABLE 11.3  SPECT System Performance (Typical Values)
                                          Parameter               Specification
                                    Number of scintillation cameras  1, 2, or 3
                                    Count sensitivity per camera  90 cps/MBq per detector
                                    (High-resolution collimator)  (200 cpm/μCi per detector)
                                    Matrix size               64 × 64; 128 × 128
                                    Pixel size                6 mm; 3 mm
                                    Spatial resolution (brain studies)  8 mm
                                    Spatial resolution (heart studies)  12 mm
                                    SPECT uniformity          15%
   345   346   347   348   349   350   351   352   353   354   355