Page 487 - Subyek Teknik Mesin - Forsthoffers Best Practice Handbook for Rotating Machinery by William E Forsthoffer
P. 487

Be st Practice 7 .36          Lube, Seal and Control Oil System Best Practices






































       Fig 7.36.5   Typical buffer gas system



       The flow on each end is then distributed back to the compressor  Preventive maintenance
       and towards the seal. This action ensures that the seal interface
       is always in contact with buffer gas, which is a controlled gas that  Referring to Figure 7.36.5, typical preventive maintenance of
       meets unit seal cleanliness requirements. Depending upon the  buffer gas systems should include the following:
       oil trap configuration, buffer gas may flow back to the unit. In
       this case, it will flow to the compressor.
          It is important to note that this system only ensures that the  Periodic valve movement
       process gas will not come in contact with the seal. Greater flow  Some buffer gas systems usually are quite stable, since differ-
       towards the compressor than towards the sour oil traps could  ential pressure of buffer gas valve and buffer gas opening can
       result in seal oil ingestion into the machine. If this is not ac-  remain relatively constant. The valve should be stroked peri-
       ceptable, a modification to this system must be made. This will  odically to ensure freedom of movement. Care must be taken to
       be discussed below.                                  ensure this action does not result in a unit trip since the com-
          Typical buffer gas velocities towards the compressor end seal  pressor seal oil or seal gas system differential pressure is refer-
       are 6e9 m/sec (20e30 ft/sec) but should take into account  enced to the buffer gas pressure.
       maximum (worn) labyrinth clearances.
                                                            Switch calibration
       Sources of buffer gas                                Periodic switch or transmitter calibration is essential to ensure
                                                            proper operation of all devices and acceptable set points.
       The sources of buffer gas can vary with the application. Buffer
       gas can be obtained directly from the compressor if the gas is
       compatible with the seal oil, or from any external source that  Flow meter measurement and calibration
       meets buffer gas specifications. When buffer gas is provided  As mentioned, the system is actually a flow control system.
       from an external source, one must consider the reliability of  Differential pressure control has historically been used because
       the source in assessing total buffer gas system reliability. Note  of its accuracy over flow control. However, there are systems
       that when buffer gas is taken from the compressor, it should  today that very accurately control flow. In order to assess the
       be superheated to a sufficient value to prevent condensation  operation of buffer gas system and the deterioration of any
       across the labyrinth. Consideration should also be given to  system components, buffer gas flow measurement is essential.
       applications that are using water and/or solvent injection to  Flow meter values should periodically be recorded by operators
       prevent fouling.                                     and flow meters should be calibrated. Flow meters should be

           458
   482   483   484   485   486   487   488   489   490   491   492