Page 254 - Electrical Safety of Low Voltage Systems
P. 254

Testing the Electrical Safety    237


                                  FIGURE 14.15
                                  Erroneous reading
                                  V 1 due to the
                                  disturbance voltage
                                  V d .




                                  V, as follows:


                                                         V = V + V                    (14.17)
                                                           1    d
                                     Figure 14.15 shows the phasor diagram of Eq. (14.17).
                                     The voltmeter then returns an erroneous result, which may inval-
                                  idate the measurement.
                                     Thus, in order to determine the correct values for touch and step
                                  potentials, a set of three measurements should be performed:

                                      1. Measurement of the r.m.s. value of V d , obtained in the absence
                                         of the test current.
                                      2. Measurement of the r.m.s. value of V 1 , by impressing the test
                                         current I t .
                                      3. Measurement of the r.m.s. value of V 2 , by impressing the test
                                         current I t of same intensity as in the previous measurement,
                                         but with inverted polarity.

                                     The measurement with inverted polarity is described by the vector
                                  diagram in Fig. 14.16.
                                     We assume that the disturbance remains constant during the two
                                  measurements, therefore V d in Figs. 14.16 and 14.17 will have the same
                                  length. However, V, although having a different angular displacement
                                  with reference to V d , will not change its magnitude even if we change
                                  the direction of the test current.



                                  FIGURE 14.16
                                  Erroneous reading V 2
                                  (test current with
                                  inverted polarity)
                                  due to the
                                  disturbance voltage
                                  V d .
   249   250   251   252   253   254   255   256   257   258   259