Page 295 - Fair, Geyer, and Okun's Water and wastewater engineering : water supply and wastewater removal
P. 295

JWCL344_ch07_230-264.qxd  8/2/10  8:44 PM  Page 255







                                                                                       7.7 Automated Optimization  255


                     EXAMPLE 7.5   MODEL CALIBRATION USING DARWIN CALIBRATOR
                                         WaterGEMS has the ability to use measured field data to calibrate a model. In many cases, data that
                                         are entered into the model are an approximation or guess. When the model results do not match
                                         field data, then parameters in the model are adjusted. Also, using the Darwin Calibrator and field
                                         data, we can locate potential differences between the real network and the model that could be
                                         caused by problems in the system (blockages, closed valves, etc.).
                                             Adjust the Hazen-Williams C factor (roughness factor) for pipes P-2, P-1, and P-8 for the pipe
                                         network from Example 7.2. During the field measurements, the tank level was 3.93 m, the pump
                                         was off, and a hydrant with a measured flow of 3,400 L/min at junction 7 (J-7) was opened to in-
                                         crease the head loss in the pipe network. The measured field data are shown below:

                                                                  Junction     Pressure (kPa)

                                                                    J-7           296.0
                                                                    J-6           406.5
                                                                    J-1           263.5
                                                                    J-2           327.0

                                         Solution:
                                             •  Open Tutorial 2 (from Example 7.2).
                                             •  From the Analysis menu, select Scenarios. Create a new base scenario named “Calibrator
                                                Analysis.”
                                             •  Click on the Calculation Options tab at the bottom of the window. Create a new Calculation
                                                Option by clicking the New button and enter “Calibrator Calculation Options” as the name.
                                                Double-click the calculation options you just created. The Time Analysis Type is Steady State.
                                             •  Go back to the Scenarios tab and right-click the Calibrator Analysis scenario and select
                                                Make Current. The red check should now be on the Calibrator Analysis scenario.
                                                Double-click on the Calibrator Analysis scenario and select the Calibrator Calculation
                                                Options in the Calculation Options field. Click the Compute  button.
                                             •  Click the Darwin Calibrator   button to create a calibration study to determine the
                                                pipe roughness factors. Click the New button to select New Calibration Study. Check to
                                                make sure the Representative Scenario window has Calibrator Analysis selected.
                                             •  Enter the field data in the Field Data Snapshots tab. Click the New button to enter new
                                                data. Enter the measured pressures on the Observed Target tab. Select the junctions
                                                from the drawing by clicking the Select button   then clicking junctions J-7 J-6, J-l,
                                                and J-2 from the drawing. Click the green check   in the Select dialog. Select Pressure
                                                (kPa) for each junction in the Attribute field. Enter the measured pressure for each
                                                junction in the Value field.
                                             •  Enter the tank level in the Boundary Overrides tab. Select the tank from the drawing by
                                                clicking the Select button then clicking the tank. Click the green check  in the Select
                                                dialog. Select Tank Level (m) in the Attribute field, then enter 3.93 m in the Value field.
                                             •  To make sure the pump is off, click the New button in the Boundary Overrides tab. Select the
                                                pump by clicking the Select button   , then clicking the pump. Click the green check   in
                                                the Select dialog. Select Pump Status in the Attribute field, then select Off in the Value field.
                                             •  Enter the additional demand at junction 7 in the Demand Adjustments tab. Select this
                                                junction from the drawing by clicking the Select button  , then clicking J-7. Click the
                                                green check  in the Select dialog. Enter 3400 L/min in the Value field.
                                             •  Select the pipes where the roughness values are to be determined in the Roughness
                                                Groups tab. Pipes with similar roughness can be grouped together, or the software can
   290   291   292   293   294   295   296   297   298   299   300