Page 231 - Instrumentation Reference Book 3E
P. 231

Microprocessor-based and intelligent flowmeters  215

              Data which are stored in the non-volatile mem-                 lntearallv mounted
             ory include:                                                    transmitter
             (1)  Transmitter  model number
             (2)  Line  size
             (3)  Material  of construction
             (4) Sensor  limits
             (5) Minimum  span
             (6)  Transmitter  software revision level.


             13.7.2  Electromagnetic flowmeters
             Since  the  first  applications  of  electromagnetic
             flowmeters in  the process  industries,  during  the
             mid-1950s, their  design has  progressed  through
             several  evolutionary  phases  and  their  perform-
             ance has now reached  a very high  standard. As
             described in Chapter 1, their principal  attributes
             are that the head loss is negligible and the output
             signal is proportional to flow rate over a very wide
             range.  The sensor is relatively insensitive to vel-
             ocity profile and hence to changes in viscosity, with
             the result  that  the  installation  requirements  are
             less severe than for most other types of flowmeter.
             Provided  that  the fluid  conductivity  is  above  a
                                                      Figure 13.1 8  Electromagnetic flowmeter with
             specified minimum value (typically 5 pS/cm), the   transmitter mounted integrally with the flowtube. Courtesy
             accuracy of measurement  is typically better than   of the lnvensys (formerly Foxboro) Company.
             1 percent. However, the system is not suitable for
             gases.                                   range value of the system to be set for a particular
               Electromagnetic  flowmeters  differ  from pres-   application.  During  power-up  and  periodically
             sure  and  many  other  transmitters  in  that  they   during normal operation. diagnostic routines are
             require considerably more power than is available   run  to  identify  and  isolate  faults  in  either  the
             in a conventional 4-2Q mA measurement  circuit.   transmitter or flowtube. Means are also provided
             However.  some  manufacturers  have  developed   to prevent false measurements when the flowtube
             transmitters  which  include  intelligent  features,   has  run  empty  and to cu:  off  the  output  signal
             and the Invensys (formerly Foxboro) 8000 Series   when the flow rate  falls below 2 percent  of  the
             (see References) is representative of them.   upper range value.
               The system involves two components: the flow-   The transmitter communicates bi-directionally
             tube which is mounted in the pipeline and senses   to a hand-held terminal (HHT), shown in Figure
             the  flow,  and  the  transmitter  which  takes  the   13.19, or to the Invensys (formerly Foxboro) I/A
             small signal developed between the electrodes by   system, using the frequency shift keying technique
             the flowing fluid and converts it into an output   with  the  signals  superimposed  on  the  4-20  mA
             signal such as 420 niA in a conventional current   measurement  loop.  The  handheld  terminal  can
             loop or a pulse output signal. The two units can   be connected at any point in a 4-20  mA measure-
             be  combined,  as  shown  in  Figure  13.18,  or   ment circuit in a general-purpose, ordinary loca-
             mounted separately.                      tion or a Division 2 area.
               The  transmitter  provides  three  electrical out-   The  information  which  can  be  displayed
             puts.  (i) An  analog signal in the range 4-20mA   includes:
             current  into a 3000 load when powered  intern-
             ally,  or  into  18009 load  when  powered  by  an   (1)  Status  of the continuous self-diagnostic rou-
             external  50V  d.c.  supply.  (ii)  A  digital  signal   tine
             which is superimposed on the 4-20  mA measure-   (2)  Day,  date, and time.
             ment loop. (iii) A low or high rate pulse output   In addition, the following information  can be
             which may  be  powered  internally  or externally.   displayed and/or configured:
             There is an LCD display which displays the flow
             rate  either  as  a  percentage  of  the  upper  range   (1)  Measurement  output (in engineering units)
             value or in engineering units.           (2)  Pulse  output parameters
               Other  features  include  a  built-in  calibrator   (3)  Flowtube  lining material
             which enables the units of measure and the upper   (4)  Flowtube  maximum flow range
   226   227   228   229   230   231   232   233   234   235   236